À retenir
In 30 women (15 pregnant, 15 non-pregnant) doing high-intensity interval running, a wearable sensor differed from a rectal probe by more than the 0.3°C limit in 47% of pregnant and 43% of non-pregnant measurements. It overestimated core temperature in pregnancy and underestimated it in non-pregnant women, from a small sample in one protocol.
Study design
Article de recherche
Reading basis
Published abstract
Reading focus
Design & applicability
Cette explication éditoriale existe en anglais, chinois simplifié et chinois traditionnel. Cette interface affiche la lecture en anglais.
Question de recherche
Some pregnant athletes continue hard exercise, and high core temperature could be a concern for the fetus, so tracking maternal temperature may matter. Yet pregnancy changes skin blood flow and heat loss, which could throw off non-invasive sensors. The authors compared the CALERA Research Sensor (CRS) with a rectal probe, the gold standard, in pregnant and non-pregnant runners.
Population et méthode
Thirty elite and recreational athletes took part: 15 pregnant (31 ± 4 years) and 15 non-pregnant (30 ± 6 years). After a separate 10-min warm-up, they ran up to five 5-min high-intensity treadmill intervals with recovery pauses, then had 20 min of post-exercise with no cool-down, at 20-21°C and 38-61% relative humidity. Agreement was assessed with Bland-Altman plots against a pre-defined ±0.3°C acceptability threshold, and differences with a paired t-test.
Résultats
The authors’ reported results. Read each estimate with its population and uncertainty.
- 1Nearly half of the measurements fell outside the ±0.3°C threshold: 47% in pregnant and 43% in non-pregnant participants, showing systematic bias.
- 2In pregnant participants the mean difference (rectal probe - CRS) was -0.23°C (95% CI: -0.36, -0.09°C), meaning CRS read higher than the rectal probe.
- 3In non-pregnant participants the mean difference was 0.16°C (95% CI: 0.03, 0.28°C), meaning CRS read lower than the rectal probe.
- 4Discrepancies were largest in the early intervals and shrank over time.
Limites et incertitude
- 1Only 15 participants per group, mixing elite and recreational athletes, from a single running protocol under 20-21°C conditions.
- 2The comparison covers one sensor and one exercise type, so it does not apply to other devices or to sport in heat.
- 3The abstract does not report results by trimester or the agreement limits beyond the mean differences.
- 4This reading is based on the abstract only.
Mise en perspective
These results suggest a skin-based core temperature sensor should not be relied on alone to judge heat exposure during hard exercise in pregnancy. Its error runs in the opposite direction from non-pregnant women. Exercise in pregnancy should be discussed with a clinician.
Vos données
If your device estimates core or skin temperature, treat it as a relative trend, not a precise reading. Compare sessions with the same device, placement and conditions, and do not use a single value to decide how hard to exercise when pregnant.
Comprendre ce type de preuve
The index does not establish a precise design. Read the methods before making causal claims; check population, comparator, outcomes and uncertainty in the source.
Résumé structuré des auteurs English
Le résumé original en anglais est reproduit avec attribution sous la licence ouverte indiquée. Il s’agit d’une source, pas d’une interprétation MoveVitals. L’analyse préparée figure ci-dessus ; vérifiez les conclusions importantes dans la source.
Background
An increasing number of pregnant women, including elite athletes, continue vigorous- and high-intensity exercise during pregnancy. Current pregnancy exercise guidelines primarily recommend at least 150 min of moderate-intensity physical activity per week and provide limited specific guidance on vigorous- and high-intensity exercise in highly trained pregnant women. High-intensity exercise can elevate core temperature (Tcore), raising concerns about maternal hyperthermia and potential fetal risks (e.g. neural tube defects, preterm birth, fetal growth restriction). Accurate monitoring of maternal Tcore may be necessary to evaluate safety. However, pregnancy-related cardiovascular and thermoregulatory adaptations, such as changes in skin blood flow and heat dissipation, may affect temperature estimates, yet few non-invasive and user-friendly methods combine accuracy with practicality during exercise. This study compared the non-invasive CALERA Research Sensor (CRS) with a rectal probe, the gold standard for Tcore measurement, during high-intensity interval running in pregnant and non-pregnant participants.
Methods
Thirty elite and recreational athletes participated, including 15 pregnant participants (31 ± 4 years) and 15 non-pregnant participants (30 ± 6 years). All completed a separate 10-min warm-up, followed by up to five 5-min high-intensity treadmill-intervals separated by recovery periods. Testing was conducted under environmental conditions of 20-21°C and 38-61% relative humidity. Analyses included the intervals, corresponding pauses, and 20 min of post-exercise, during which no cool-down was performed. Agreement was assessed using Bland-Altman plots, with a pre-defined ±0.3°C acceptability threshold. Differences were analyzed using paired t -test.
Results
Nearly half of the measurements exceeded the predefined threshold (47% in pregnant vs. 43% in non-pregnant participants), reflecting systematic bias between methods. In pregnant participants, the mean difference (rectal probe - CRS) was -0.23°C (95% CI: -0.36, -0.09°C), indicating that CRS overestimated Tcore relative to the rectal probe. In contrast, in non-pregnant participants, the mean difference was 0.16°C (95% CI: 0.03, 0.28°C), indicating that CRS underestimated Tcore. Discrepancies were most pronounced during the early intervals but diminished over time.
Conclusion
This study shows opposing systematic bias of CRS in pregnant and non-pregnant women compared with Tcore measured using a rectal probe, limiting its use as a stand-alone tool in pregnancy and highlighting the need for non-invasive temperature sensor development.
Sources et attribution
Comparison of a wearable core temperature sensor with rectal thermometry during high-intensity interval running in pregnant and non-pregnant women.
Brevik-Persson S, Melau J, Dalhaug EM, Gjestvang C, Haakstad LAH.
Journal of exercise science and fitness · 2026-10-01
DOI: 10.1016/j.jesf.2026.200529 ↗PubMed · 42921436 ↗Europe PMC · Licence indiquée: cc by ↗Abstract © the original authors, reproduced under the recorded open licence. Translations support reading; use the English record to check terminology and numerical results. Our methodological interpretation is separate from the authors’ conclusions.
Interprétation éditoriale MoveVitals · contenu éducatif · aucune évaluation clinique individuelle enregistrée
Sources vérifiées 2026-11-01
Guides pratiques
À lire ensemble
Activity–stroke and heart attack curves differ by accelerometer processing method
Among 90,237 UK Biobank adults free of cardiovascular disease at baseline, followed until December 2022 for 1,298 strokes and 2,031 myocardial infarctions, the shape of the link between moderate-vigorous physical activity (MVPA) and each outcome depended on how the accelerometer data were processed. For stroke, hazard fell in a straight line only with the machine-learning metric; for myocardial infarction, two metrics gave a curved decline and activity counts a straight one. The abstract gives no hazard ratios, so it shows that methods disagree, not which one is right, and the findings are observational associations.
Preventive Medicine
Smartwatch data as forensic evidence: heart rate is best validated, the rest uneven
A systematic review and meta-analysis of studies on smartwatches from four major brands, framed around digital forensics, found that heart rate was both the most studied and the most accurate measure. Step count, distance, energy expenditure, activity classification and sleep were less consistently validated, and results varied by brand, protocol and reporting. The abstract gives no pooled numbers, and most studies were run in healthy people in laboratories, so it cannot say how a particular watch performs for a particular person in daily life.
Forensic Science International
Activity trackers: about 1,800 more steps a day, but small effects on other outcomes
An umbrella review of 39 systematic reviews and meta-analyses, covering 163,992 participants, found that activity trackers were associated with more physical activity (SMD 0.3–0.6), equating to about 1,800 extra steps and 40 minutes more walking per day, and with a bodyweight reduction of about 1 kg. Effects on blood pressure, cholesterol, glycosylated haemoglobin, quality of life and pain were typically small and often non-significant. The abstract does not say what else came with the trackers in these programs or how much the reviews overlap, so the gains cannot be credited to the device alone.
The Lancet Digital Health
Seven sleep trackers: good at spotting sleep, weak on wake and sleep stages
In a sleep laboratory, 34 healthy young adults wore a subset of seven consumer sleep trackers for three nights. Against polysomnography every device detected sleep very well (sensitivity 0.93 or higher) but caught wake poorly (specificity 0.18–0.54), and five of the seven matched or beat research actigraphy (specificity 0.39). Sleep-stage estimates were inconsistent, and the participants were healthy adults aged 18 to 35 in a fixed 8-hour bed window, with each device tested in only 11 to 20 of them, so the results say little about people with sleep problems.
Sleep

