穿戴裝置與測量科學證據回顧4 分鐘閱讀

脈搏血氧儀各種膚色下讀數都略偏高,穿戴式脈率則未見膚色偏差

MoveVitals 編輯部 · 研究發表 · 解讀更新

原文:Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis. · Journal of medical Internet research · 2024 · 論文原文 ↗

先看重點

在23項脈搏血氧飽和度研究(n = 59,684)和4項穿戴式裝置脈率研究(n = 176)中,淺色、中等和深色皮膚的準確度均未達到監管閾值,深色皮膚的SpO2平均偏差最高(1.27%,95%置信區間0.58至1.95)。納入的穿戴式裝置研究很少,且深色皮膚的脈率誤差範圍最寬。

證據類型

證據回顧

解讀依據

公開摘要

閱讀重點

方法與適用範圍

研究在問什麼

脈搏血氧儀和穿戴式裝置透過皮膚透光(光電容積描記,PPG)來估算血氧飽和度(SpO2)和脈率。如果膚色會影響讀數,某些族群得到的醫療服務可能就不夠可靠。研究者對照監管標準,按膚色檢查了準確度和偏差。

研究對象與方法

對23項脈搏血氧飽和度研究(n = 59,684;197,353對SpO2與動脈血配對觀察)和4項穿戴式裝置脈率研究(n = 176;140,771對PPG與心電圖配對觀察)進行系統評價和統合分析。依據美國國家標準協會、美國醫療器械促進協會和國際電工委員會的標準,將SpO2準確度均方根值與3%的監管閾值比較,將脈率95%一致性界限與+5或-5次/分比較。按淺色、中等和深色色素沉著,用平均偏差及95%置信區間評估偏差和臨床相關性。

主要發現

以下為作者報告的結果。數值須連同族群和不確定性一起閱讀。

  1. 1
    SpO2準確度均方根值為:淺色3.96%,中等4.71%,深色4.15%,均高於3%的閾值。
  2. 2
    SpO2合併平均偏差:淺色0.70%(95%置信區間0.17%至1.22%),中等0.27%(95%置信區間-0.64%至1.19%),深色1.27%(95%置信區間0.58%至1.95%)。淺色和深色皮膚的讀數被顯著高估,但這可能沒有臨床意義。
  3. 3
    穿戴式裝置脈率的95%一致性界限:淺色-16.02至13.54,中等-18.62至16.84,深色-33.69至32.54次/分,均寬於+5或-5次/分。
  4. 4
    脈率合併偏差:淺色-1.24次/分(95%置信區間-5.31至2.83),中等-0.89次/分(95%置信區間-3.70至1.93),深色-0.57次/分(95%置信區間-9.44至8.29),不同膚色之間沒有統計學或臨床意義上的顯著偏差。

限制與不確定性

  1. 1
    脈率資料僅來自4項穿戴式裝置研究、176人,深色皮膚的區間非常寬(-9.44至8.29次/分),不能排除存在較小差異。
  2. 2
    摘要沒有說明各研究如何劃分膚色(不同量表和評估者可能不同),也沒有給出每個膚色組的研究數量。
  3. 3
    平均值會掩蓋最關鍵的情形:平均偏差接近零的同時仍可能有很大的個體誤差,寬的一致性界限正說明了這一點。
  4. 4
    以上解讀僅基於摘要。

日常健康啟示

無論膚色深淺,脈搏血氧儀和穿戴式裝置的脈率讀數都可能超出監管可接受的誤差,所以單個數字並不是精確測量。本分析中穿戴式裝置的平均脈率偏差與膚色無關,但深色皮膚的個體讀數波動更大。如果你感覺不適,不要因為讀數看起來正常就放心,應尋求醫療建議。

如何連結自己的資料

把穿戴式裝置的脈率或SpO2讀數當作同一裝置、佩戴方式一致(貼合、靜止、皮膚乾淨乾燥)下的趨勢,而不是精確數值。請比較夜間的整體規律而不是單次抽查,對異常讀數應用醫用級裝置或請醫師確認。

如何理解這類證據

先看納入了哪些設計、文獻搜尋與偏誤評估如何完成,以及搜尋截止日期;發表新不代表底層證據都是最新的。

作者的結構化摘要 English

以下英文摘要依所記錄的開放授權轉載,屬於作者原始材料,並非 MoveVitals 解讀。上方已提供研究解讀,分別說明作者結果與本站解釋;重要結論可直接查核原文。

Background

Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO_(2)) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities.

Objective

This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO_(2) and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO_(2) and PR according to skin pigmentation.

Methods

We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO_(2)-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO_(2) accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or -5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI.

Results

For SpO_(2), accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI -0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from -16.02 to 13.54, from -18.62 to 16.84, and from -33.69 to 32.54, and pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm, -0.89 (95% CI -3.70 to 1.93) bpm, and -0.57 (95% CI -9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.

Conclusions

SpO_(2) and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO_(2) for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.

來源與署名

Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.

Singh S, Bennett MR, Chen C, Shin S, Ghanbari H, Nelson BW.

Journal of medical Internet research · 2024-10-10

DOI: 10.2196/62769 ↗PubMed · 39388258 ↗Europe PMC · 所記錄的開放授權: cc by ↗

摘要著作權歸原作者,依來源紀錄的開放授權轉載。中文譯文協助閱讀,英文摘要供查核術語及具體數值;本站的方法學解釋與作者結論分別呈現。

MoveVitals 編輯解讀 · 健康教育內容 · 未記錄個人臨床審閱

來源查核日期 2026-11-01

實用健康指南

放在一起讀

·世代研究

加速度計處理方法不同,身體活動與中風、心肌梗死的關聯曲線也不同

在英國生物銀行中,基線無心血管疾病的90237名成年人追蹤至2022年12月,共發生1298例中風和2031例心肌梗死;中高強度身體活動(MVPA)與這兩種結局的關聯曲線形狀,取決於加速度計資料的處理方法。就中風而言,只有機器學習法得到直線下降的風險;就心肌梗死而言,機器學習法和LFENMO法呈曲線下降,活動計數法呈直線下降。摘要沒有給出風險比,只能說明不同方法的結論不一致,不能說明哪種方法更正確,而且這些只是觀察性關聯。

Preventive Medicine

重點論文精講閱讀論文 →
·測量驗證研究/回顧

智慧手錶資料能當「證人」嗎:心率驗證得最充分,其餘證據參差

一項圍繞數字取證展開的系統評價和統合分析,梳理了四大主流品牌智慧手錶的驗證研究,發現心率是研究最多、也最準確的指標。步數、距離、能量消耗、活動分類和睡眠的驗證不夠一致,結果因品牌、方案和報告方式而異。摘要沒有給出彙總資料,且多數研究是在健康族群中、於實驗室內完成的,所以無法說明某一塊手錶對某個人在日常生活中的表現。

Forensic Science International

重點論文精講閱讀論文 →
·回顧的回顧

活動追蹤器:每天多走約1800步,對其他健康指標影響較小

這項傘形綜述彙總了39篇系統評價和統合分析,涉及163992名參與者;結果顯示使用活動追蹤器與身體活動增加相關(標準化均數差SMD為0.3–0.6),相當於每天多走約1800步、多步行40分鐘,體重減輕約1公斤。對血壓、膽固醇、糖化血紅蛋白、生活質量和疼痛的影響通常很小,且多數不顯著。摘要沒有說明這些專案裡追蹤器還搭配了什麼,也沒有說明各評價之間重疊多少,因此不能把效果完全歸功於裝置本身。

The Lancet Digital Health

重點論文精講閱讀論文 →
·測量驗證研究/回顧

七款睡眠追蹤裝置:識別睡著很準,識別醒著和睡眠分期較差

在睡眠實驗室裡,34名健康的年輕成年人連續三晚佩戴七款消費級睡眠追蹤裝置中的若干款。與多導睡眠監測相比,所有裝置識別睡眠都很準(靈敏度均不低於0.93),但識別清醒較差(特異度0.18–0.54),七款中有五款與研究用活動記錄儀(特異度0.39)持平或更好;睡眠分期的估計結果不穩定。受試者是18至35歲的健康成年人,臥床時間固定為8小時,每款裝置只在其中11至20人身上測試,因此對有睡眠問題的人幾乎沒有參考價值。

Sleep

重點論文精講閱讀論文 →
穿戴裝置與測量科學 →