可穿戴与测量科学测量验证研究/综述3 分钟阅读

一人24小时实测:Apple Watch 3与Fitbit Charge 2心率误差在10%以内

MoveVitals 编辑部 · 研究发表 · 解读更新

原文:Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study. · JMIR mHealth and uHealth · 2019 · 论文原文 ↗

先看重点

在1名受试者的24小时内,Apple Watch 3与动态心电图平均相差-1.80次/分,Fitbit Charge 2相差-3.47次/分,两者平均绝对误差都接近6%。样本只有1人(第一作者)和2款设备,只能作初步参考,不能推广。

证据类型

测量验证研究/综述

解读依据

公开摘要

阅读重点

参照与测量误差

研究在问什么

可穿戴设备的心率准确性通常在实验室的跑步机或功率车上测试,而不是在有家务和睡眠的普通一天里。作者想检验两款常见的腕戴设备在连续24小时的真实使用中,与金标准心电图的接近程度。

研究了谁,如何研究

单受试者验证研究:1名受试者(第一作者,29岁白种男性)同时佩戴Apple Watch 3、Fitbit Charge 2和作为参照的动态心电图(Vrije Universiteit Ambulatory Monitoring System)24小时,共采集102,740次心跳。数据按5种情境划分:静坐、步行、跑步、日常活动(如做家务、刷牙)和睡眠。准确性用平均误差、平均绝对误差和平均绝对百分比误差评价,并用Bland-Altman分析和一致性相关系数评估一致性。

主要发现

以下为作者报告的结果。数字须连同人群和不确定性一起阅读。

  1. 1
    Apple Watch 3与心电图相比,24小时内平均差值为-1.80次/分,平均绝对误差为5.86%,平均一致性为95%。
  2. 2
    Fitbit Charge 2与心电图相比,24小时内平均差值为-3.47次/分,平均绝对误差为5.96%,平均一致性为91%。
  3. 3
    按作者的标准,两款设备在24小时整体及各情境下的误差都在<±10%的可接受范围内,只有Apple Watch 3在日常活动情境下例外。
  4. 4
    结果因情境而异,所以24小时平均表现良好,并不意味着每种活动都同样准确。

局限与不确定性

  1. 1
    只有1名受试者佩戴,且他是第一作者,结果不一定适用于其他肤色、腕围、佩戴松紧或心律的人。
  2. 2
    只测试了2款较早的型号(Apple Watch 3、Fitbit Charge 2),新款设备和新版固件不在其中。
  3. 3
    只记录了一个24小时,仅反映当天的情境;作者自己也称结果为初步结论。
  4. 4
    以上解读仅基于摘要,摘要没有给出各情境的具体数值。

对日常健康的启示

结果提示,对这名受试者来说,这两款设备的腕部心率在普通一天里与心电图较为接近,支持把它们用于研究和日常监测。但这不代表你手上的设备同样准确,也不支持用可穿戴设备心率来做诊断。

如何联系自己的数据

手表测得的静息心率和日常心率,更适合在同一设备、同样的佩戴方式(贴紧、同一只手腕)下看趋势。做家务等手臂活动不规律时误差可能更大,单次读数要谨慎对待。

如何理解这类证据

一致性取决于设备、固件、参照、人群和场景。相关性不等于一致性;验证某个指标不等于已证明诊断或治疗获益。

作者的结构化摘要 English

以下英文摘要按记录的开放许可转载,属于作者原始材料,不是 MoveVitals 的解读。上方已提供研究解读,分别说明作者结果与本站解释;重要结论可直接核对原文。

Background

Wrist-worn smart watches and fitness monitors (ie, wearables) have become widely adopted by consumers and are gaining increased attention from researchers for their potential contribution to naturalistic digital measurement of health in a scalable, mobile, and unobtrusive way. Various studies have examined the accuracy of these devices in controlled laboratory settings (eg, treadmill and stationary bike); however, no studies have investigated the heart rate accuracy of wearables during a continuous and ecologically valid 24-hour period of actual consumer device use conditions.

Objective

The aim of this study was to determine the heart rate accuracy of 2 popular wearable devices, the Apple Watch 3 and Fitbit Charge 2, as compared with the gold standard reference method, an ambulatory electrocardiogram (ECG), during consumer device use conditions in an individual. Data were collected across 5 daily conditions, including sitting, walking, running, activities of daily living (ADL; eg, chores, brushing teeth), and sleeping.

Methods

One participant, (first author; 29-year-old Caucasian male) completed a 24-hour ecologically valid protocol by wearing 2 popular wrist wearable devices (Apple Watch 3 and Fitbit Charge 2). In addition, an ambulatory ECG (Vrije Universiteit Ambulatory Monitoring System) was used as the gold standard reference method, which resulted in the collection of 102,740 individual heartbeats. A single-subject design was used to keep all variables constant except for wearable devices while providing a rapid response design to provide initial assessment of wearable accuracy for allowing the research cycle to keep pace with technological advancements. Accuracy of these devices compared with the gold standard ECG was assessed using mean error, mean absolute error, and mean absolute percent error. These data were supplemented with Bland-Altman analyses and concordance class correlation to assess agreement between devices.

Results

The Apple Watch 3 and Fitbit Charge 2 were generally highly accurate across the 24-hour condition. Specifically, the Apple Watch 3 had a mean difference of -1.80 beats per minute (bpm), a mean absolute error percent of 5.86%, and a mean agreement of 95% when compared with the ECG across 24 hours. The Fitbit Charge 2 had a mean difference of -3.47 bpm, a mean absolute error of 5.96%, and a mean agreement of 91% when compared with the ECG across 24 hours. These findings varied by condition.

Conclusions

The Apple Watch 3 and the Fitbit Charge 2 provided acceptable heart rate accuracy (<±10%) across the 24 hour and during each activity, except for the Apple Watch 3 during the daily activities condition. Overall, these findings provide preliminary support that these devices appear to be useful for implementing ambulatory measurement of cardiac activity in research studies, especially those where the specific advantages of these methods (eg, scalability, low participant burden) are particularly suited to the population or research question.

来源与署名

Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study.

Nelson BW, Allen NB.

JMIR mHealth and uHealth · 2019-03-11

DOI: 10.2196/10828 ↗PubMed · 30855232 ↗Europe PMC · 记录的开放许可: cc by ↗

摘要版权归原作者,依据来源记录的开放许可转载。中文译文辅助阅读,英文摘要用于核对术语与具体数值;本站的方法学解释与作者结论分别呈现。

MoveVitals 编辑解读 · 健康教育内容 · 未记录个人临床审阅

来源核验日期 2026-11-01

实用健康指南

放在一起读

·队列研究

加速度计处理方法不同,身体活动与卒中、心肌梗死的关联曲线也不同

在英国生物银行中,基线无心血管疾病的90237名成年人随访至2022年12月,共发生1298例卒中和2031例心肌梗死;中高强度身体活动(MVPA)与这两种结局的关联曲线形状,取决于加速度计数据的处理方法。就卒中而言,只有机器学习法得到直线下降的风险;就心肌梗死而言,机器学习法和LFENMO法呈曲线下降,活动计数法呈直线下降。摘要没有给出风险比,只能说明不同方法的结论不一致,不能说明哪种方法更正确,而且这些只是观察性关联。

Preventive Medicine

重点论文精讲阅读论文 →
·测量验证研究/综述

智能手表数据能当“证人”吗:心率验证得最充分,其余证据参差

一项围绕数字取证展开的系统评价和荟萃分析,梳理了四大主流品牌智能手表的验证研究,发现心率是研究最多、也最准确的指标。步数、距离、能量消耗、活动分类和睡眠的验证不够一致,结果因品牌、方案和报告方式而异。摘要没有给出汇总数据,且多数研究是在健康人群中、于实验室内完成的,所以无法说明某一块手表对某个人在日常生活中的表现。

Forensic Science International

重点论文精讲阅读论文 →
·综述的综述

活动追踪器:每天多走约1800步,对其他健康指标影响较小

这项伞形综述汇总了39篇系统评价和荟萃分析,涉及163992名参与者;结果显示使用活动追踪器与身体活动增加相关(标准化均数差SMD为0.3–0.6),相当于每天多走约1800步、多步行40分钟,体重减轻约1公斤。对血压、胆固醇、糖化血红蛋白、生活质量和疼痛的影响通常很小,且多数不显著。摘要没有说明这些项目里追踪器还搭配了什么,也没有说明各评价之间重叠多少,因此不能把效果完全归功于设备本身。

The Lancet Digital Health

重点论文精讲阅读论文 →
·测量验证研究/综述

七款睡眠追踪设备:识别睡着很准,识别醒着和睡眠分期较差

在睡眠实验室里,34名健康的年轻成年人连续三晚佩戴七款消费级睡眠追踪设备中的若干款。与多导睡眠监测相比,所有设备识别睡眠都很准(灵敏度均不低于0.93),但识别清醒较差(特异度0.18–0.54),七款中有五款与研究用活动记录仪(特异度0.39)持平或更好;睡眠分期的估计结果不稳定。受试者是18至35岁的健康成年人,卧床时间固定为8小时,每款设备只在其中11至20人身上测试,因此对有睡眠问题的人几乎没有参考价值。

Sleep

重点论文精讲阅读论文 →
可穿戴与测量科学 →