Pulse oximeters read slightly high across skin tones; wearable pulse rate shows no skin-tone bias

Colectivo editorial MoveVitals · Study published · Reading updated

Original paper:Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis. · Journal of medical Internet research · 2024 · Artículo original ↗

Idea principal

Across 23 pulse oximetry studies (n = 59,684) and 4 wearable pulse rate studies (n = 176), accuracy fell short of regulatory thresholds for light, medium and dark skin alike, with SpO2 mean bias highest in dark pigmentation (1.27%, 95% CI 0.58 to 1.95). Few wearable studies are included, and wearable pulse rate errors were widest for dark skin.

Study design

Revisión de evidencia

Reading basis

Published abstract

Reading focus

Design & applicability

Esta explicación editorial está disponible en inglés, chino simplificado y chino tradicional. Esta interfaz muestra la lectura en inglés.

Pregunta de investigación

Pulse oximeters and wearables use light through the skin (photoplethysmography, PPG) to estimate blood oxygen (SpO2) and pulse rate. If skin pigmentation affects the readings, some groups may get less reliable care. The authors checked accuracy and bias by skin pigmentation against regulatory standards.

Población y método

Systematic review and meta-analysis of 23 pulse oximetry studies (n = 59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable pulse rate studies (n = 176; 140,771 paired PPG-electrocardiography observations). SpO2 accuracy root-mean-square values were compared with the regulatory threshold of 3%, and pulse rate 95% limits of agreement with +5 or -5 beats per minute (bpm), following American National Standards Institute, Association for the Advancement of Medical Instrumentation and International Electrotechnical Commission standards. Bias and clinical relevance were evaluated using mean bias and 95% CI, for light, medium and dark pigmentation.

Resultados

The authors’ reported results. Read each estimate with its population and uncertainty.

  1. 1
    SpO2 accuracy root-mean-square values were 3.96% (light), 4.71% (medium) and 4.15% (dark), all above the 3% threshold.
  2. 2
    Pooled SpO2 mean biases were 0.70% (95% CI 0.17% to 1.22%) for light, 0.27% (95% CI -0.64% to 1.19%) for medium and 1.27% (95% CI 0.58% to 1.95%) for dark pigmentation. Readings were significantly overestimated for light and dark skin, but this may not be clinically relevant.
  3. 3
    Wearable pulse rate 95% limits of agreement were -16.02 to 13.54 (light), -18.62 to 16.84 (medium) and -33.69 to 32.54 bpm (dark), all wider than +5 or -5 bpm.
  4. 4
    Pooled pulse rate biases were -1.24 bpm (95% CI -5.31 to 2.83), -0.89 bpm (95% CI -3.70 to 1.93) and -0.57 bpm (95% CI -9.44 to 8.29) for light, medium and dark pigmentation, with no statistically or clinically significant bias by skin tone.

Límites e incertidumbre

  1. 1
    Only 4 wearable studies with 176 people contributed pulse rate data, and the very wide interval for dark pigmentation (-9.44 to 8.29 bpm) means that small differences cannot be ruled out.
  2. 2
    The abstract does not say how skin pigmentation was classified across studies, which can differ between scales and examiners, and it does not give how many studies contributed to each group.
  3. 3
    Averages hide the cases that matter most: bias near zero can coexist with large individual errors, as the wide limits of agreement show.
  4. 4
    This reading is based on the abstract only.

Relevancia práctica

Pulse oximeters and wearable pulse rate readings can be off by more than regulators accept, for people of every skin tone, so a single number is not a precise measurement. In this analysis the average pulse rate bias from wearables did not depend on skin tone, but individual readings were noisier for darker skin. If you feel unwell, do not rely on a reading that looks normal; seek medical advice.

Tus datos

Use a wearable pulse rate or SpO2 reading as a trend from the same device, worn the same way (snug, still, on clean dry skin), rather than as an exact value. Compare overnight patterns rather than single spot checks, and confirm unusual readings with a medical-grade device or a clinician.

Cómo interpretar la evidencia

Check which designs were included, how searches and bias assessment were conducted, and the last search date. Publication date alone does not make the underlying evidence current.

Resumen estructurado de los autores English

El resumen original en inglés se reproduce con atribución bajo la licencia indicada. Es una fuente, no una interpretación de MoveVitals. El análisis preparado figura arriba; verifica los resultados importantes en la fuente.

Background

Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO_(2)) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities.

Objective

This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO_(2) and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO_(2) and PR according to skin pigmentation.

Methods

We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO_(2)-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO_(2) accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or -5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI.

Results

For SpO_(2), accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI -0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from -16.02 to 13.54, from -18.62 to 16.84, and from -33.69 to 32.54, and pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm, -0.89 (95% CI -3.70 to 1.93) bpm, and -0.57 (95% CI -9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.

Conclusions

SpO_(2) and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO_(2) for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.

Fuentes y atribución

Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.

Singh S, Bennett MR, Chen C, Shin S, Ghanbari H, Nelson BW.

Journal of medical Internet research · 2024-10-10

DOI: 10.2196/62769 ↗PubMed · 39388258 ↗Europe PMC · Licencia registrada: cc by ↗

Abstract © the original authors, reproduced under the recorded open licence. Translations support reading; use the English record to check terminology and numerical results. Our methodological interpretation is separate from the authors’ conclusions.

Interpretación editorial de MoveVitals · contenido educativo · sin revisión clínica individual registrada

Fuentes verificadas 2026-11-01

Guías prácticas

Leer en contexto

·Estudio de cohorte

Activity–stroke and heart attack curves differ by accelerometer processing method

Among 90,237 UK Biobank adults free of cardiovascular disease at baseline, followed until December 2022 for 1,298 strokes and 2,031 myocardial infarctions, the shape of the link between moderate-vigorous physical activity (MVPA) and each outcome depended on how the accelerometer data were processed. For stroke, hazard fell in a straight line only with the machine-learning metric; for myocardial infarction, two metrics gave a curved decline and activity counts a straight one. The abstract gives no hazard ratios, so it shows that methods disagree, not which one is right, and the findings are observational associations.

Preventive Medicine

Explicaciones detalladasLeer estudio →
·Estudio/revisión de validación

Smartwatch data as forensic evidence: heart rate is best validated, the rest uneven

A systematic review and meta-analysis of studies on smartwatches from four major brands, framed around digital forensics, found that heart rate was both the most studied and the most accurate measure. Step count, distance, energy expenditure, activity classification and sleep were less consistently validated, and results varied by brand, protocol and reporting. The abstract gives no pooled numbers, and most studies were run in healthy people in laboratories, so it cannot say how a particular watch performs for a particular person in daily life.

Forensic Science International

Explicaciones detalladasLeer estudio →
·Revisión de revisiones

Activity trackers: about 1,800 more steps a day, but small effects on other outcomes

An umbrella review of 39 systematic reviews and meta-analyses, covering 163,992 participants, found that activity trackers were associated with more physical activity (SMD 0.3–0.6), equating to about 1,800 extra steps and 40 minutes more walking per day, and with a bodyweight reduction of about 1 kg. Effects on blood pressure, cholesterol, glycosylated haemoglobin, quality of life and pain were typically small and often non-significant. The abstract does not say what else came with the trackers in these programs or how much the reviews overlap, so the gains cannot be credited to the device alone.

The Lancet Digital Health

Explicaciones detalladasLeer estudio →
·Estudio/revisión de validación

Seven sleep trackers: good at spotting sleep, weak on wake and sleep stages

In a sleep laboratory, 34 healthy young adults wore a subset of seven consumer sleep trackers for three nights. Against polysomnography every device detected sleep very well (sensitivity 0.93 or higher) but caught wake poorly (specificity 0.18–0.54), and five of the seven matched or beat research actigraphy (specificity 0.39). Sleep-stage estimates were inconsistent, and the participants were healthy adults aged 18 to 35 in a fixed 8-hour bed window, with each device tested in only 11 to 20 of them, so the results say little about people with sleep problems.

Sleep

Explicaciones detalladasLeer estudio →
Dispositivos y medición →