Idea principal
Over 24 hours in 28 people, a Samsung Gear Sport smartwatch gave acceptable heart rate and most heart rate variability measures during sleep compared with a chest ECG monitor. While awake only average heart rate and AVNN (average beat-to-beat interval) were reasonably accurate, and the other HRV measures had high errors.
Study design
Estudio/revisión de validación
Reading basis
Published abstract
Reading focus
Reference & error
Esta explicación editorial está disponible en inglés, chino simplificado y chino tradicional. Esta interfaz muestra la lectura en inglés.
Pregunta de investigación
Smartwatches estimate heart rate and pulse rate variability, a stand-in for HRV, from light reflected off the skin (photoplethysmography, PPG). That signal is easily disturbed by movement and noise, and few studies test it in everyday life. The authors compared a Samsung Gear Sport against a medical-grade chest ECG over a normal day.
Población y método
Validation study. Twenty-eight participants (14 male, 14 female) wore a Samsung Gear Sport and a Shimmer3 ECG device for 24 hours while following daily routines. Heart rate and time-domain and frequency-domain HRV measures (including AVNN, SDNN, RMSSD, pNN50, LF, HF and LF/HF) were analyzed separately for sleep and awake time. The watch was compared against the ECG using Pearson correlation, Bland-Altman plots and linear regression.
Resultados
The authors’ reported results. Read each estimate with its population and uncertainty.
- 1During sleep, the watch and ECG correlated highly for heart rate, time-domain HRV, LF and HF, and moderately for LF/HF. Mean biases were low for heart rate, time-domain HRV and LF/HF, and moderate for LF and HF.
- 2During sleep, regression error variance was low for heart rate, AVNN and pNN50, moderate for SDNN, RMSSD, LF and HF, and high for LF/HF.
- 3While awake, correlation was high for AVNN, moderate for heart rate and low for the other parameters. RMSSD and SDNN had low mean bias; the rest had moderate bias.
- 4While awake, only AVNN had moderate error variance; all other parameters had high error variance.
Límites e incertidumbre
- 1The sample was 28 people, and the abstract gives no ages, health status or how much people moved, so results may not fit everyone.
- 2Only one watch model was tested (the Gear Sport), so other brands or newer models may perform differently.
- 3The abstract describes accuracy in words such as low, moderate and high rather than giving numerical limits of agreement.
- 4Agreement during sleep does not mean the watch can detect health problems; the study did not test that.
Relevancia práctica
This supports trusting a wrist device more for heart rate and for overnight HRV than for daytime HRV snapshots. It does not show that any watch's readings are clinically accurate, and it should not be used to decide about heart conditions, which belong with your clinician.
Tus datos
If you track HRV, use the overnight or sleep-time value from the same device rather than random daytime readings. Compare your own multi-week trend, not a single night, and don't compare numbers across different brands.
Cómo interpretar la evidencia
Agreement is specific to the device, firmware, reference, population and setting. Correlation is not agreement, and validating a metric is not proof of diagnostic or treatment benefit.
Resumen estructurado de los autores English
El resumen original en inglés se reproduce con atribución bajo la licencia indicada. Es una fuente, no una interpretación de MoveVitals. El análisis preparado figura arriba; verifica los resultados importantes en la fuente.
Background
Photoplethysmography (PPG) is a low-cost and easy-to-implement method to measure vital signs, including heart rate (HR) and pulse rate variability (PRV) which widely used as a substitute of heart rate variability (HRV). The method is used in various wearable devices. For example, Samsung smartwatches are PPG-based open-source wristbands used in remote well-being monitoring and fitness applications. However, PPG is highly susceptible to motion artifacts and environmental noise. A validation study is required to investigate the accuracy of PPG-based wearable devices in free-living conditions.
Objective
We evaluate the accuracy of PPG signals-collected by the Samsung Gear Sport smartwatch in free-living conditions-in terms of HR and time-domain and frequency-domain HRV parameters against a medical-grade chest electrocardiogram (ECG) monitor.
Methods
We conducted 24-hours monitoring using a Samsung Gear Sport smartwatch and a Shimmer3 ECG device. The monitoring included 28 participants (14 male and 14 female), where they engaged in their daily routines. We evaluated HR and HRV parameters during the sleep and awake time. The parameters extracted from the smartwatch were compared against the ECG reference. For the comparison, we employed the Pearson correlation coefficient, Bland-Altman plot, and linear regression methods.
Results
We found a significantly high positive correlation between the smartwatch's and Shimmer ECG's HR, time-domain HRV, LF, and HF and a significant moderate positive correlation between the smartwatch's and shimmer ECG's LF/HF during sleep time. The mean biases of HR, time-domain HRV, and LF/HF were low, while the biases of LF and HF were moderate during sleep. The regression analysis showed low error variances of HR, AVNN, and pNN50, moderate error variances of SDNN, RMSSD, LF, and HF, and high error variances of LF/HF during sleep. During the awake time, there was a significantly high positive correlation of AVNN and a moderate positive correlation of HR, while the other parameters indicated significantly low positive correlations. RMSSD and SDNN showed low mean biases, and the other parameters had moderate mean biases. In addition, AVNN had moderate error variance while the other parameters indicated high error variances.
Conclusion
The Samsung smartwatch provides acceptable HR, time-domain HRV, LF, and HF parameters during sleep time. In contrast, during the awake time, AVNN and HR show satisfactory accuracy, and the other HRV parameters have high errors.
Fuentes y atribución
A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability.
Sarhaddi F, Kazemi K, Azimi I, Cao R, Niela-Vilén H, Axelin A, Liljeberg P, Rahmani AM.
PloS one · 2022-12-08
DOI: 10.1371/journal.pone.0268361 ↗PubMed · 36480505 ↗Europe PMC · Licencia registrada: cc by ↗Abstract © the original authors, reproduced under the recorded open licence. Translations support reading; use the English record to check terminology and numerical results. Our methodological interpretation is separate from the authors’ conclusions.
Interpretación editorial de MoveVitals · contenido educativo · sin revisión clínica individual registrada
Fuentes verificadas 2026-11-01
Guías prácticas
Leer en contexto
Short-term HRV was higher during slow breathing in 962 military men, injured or not
In 962 military men (median age 37.6 years), every short-term heart rate variability (HRV) measure was higher during slow breathing at six breaths a minute than during spontaneous breathing, injured or not; in fast resting breathers RMSSD rose from 15.57 to 21.89 ms. It was one session in a fixed order, so it shows an immediate effect, not lasting benefit.
European Journal of Applied Physiology
Cold-water immersion: no strength gain, briefly weaker jumps, soreness relief uncertain
Across 22 randomised trials (375 participants, mostly men), one cold-water immersion after exercise did not change maximal isometric strength (g 0.08, 95% CI −0.11 to 0.26) and was followed by weaker jumping straight afterwards (g −0.68), gone by 24–48 hours. Lower soreness and creatine kinase looked favourable but weakened in sensitivity checks, and certainty was low to very low.
PeerJ
Cold-water immersion after hard exercise: less soreness, better power, no strength gain
In a meta-analysis of 52 randomised trials, cold-water immersion after high-intensity exercise was followed 24 hours later by lower soreness (SMD −0.89), higher perceived recovery (0.66) and better muscular power (0.22), with no effect on strength. Participants could not be blinded, most were men, and many pooled results rest on few studies with high heterogeneity.
Sports Medicine
Acute:chronic workload ratio: no evidence it can reduce injury risk, authors argue
The authors argue there is no evidence for using the acute:chronic workload ratio (ACWR) in training-load management or to reduce injury risk: no study has properly estimated a causal effect, and the ratio is statistically inaccurate and ambiguous. This is a methodological critique, not a new data analysis, so it does not show that training load is unrelated to injury.
International Journal of Sports Physiology and Performance